Power
|
|
100 to 240 V ± 10 %, 50/60 Hz
|
Power consumption (rated)
|
|
0.8-0.5 A
|
Dimensions (W×H×D)
|
|
390×110×485 mm (incl.projection)
|
Weight
|
|
9.0 Kg
|
Observation
|
Type of color
|
NTSC/PAL
|
|
Digital HDTV
|
HD-SDI: 2, DVI-D: 2
|
|
Analog/Digital HDTV
|
DVI-I: 1
|
|
Analog SDTV
|
RGB TV: 1, S VIDEO: 1, VIDEO: 1
|
|
Screen resolution
|
SXGA (Default), Full HD
|
|
Color adjustment
|
Brightness, Red, Green, Blue, Red tone, Chroma in nine levels (-4 to +4). Contrast in five levels (-1 to +4)
|
|
Contrast
|
Available in three levels (-1 to +1).
|
|
Iris mode
|
Function to control the screen brightness. AVE (controls brightness in general), PEAK (controls brightness in highlight areas), AUTO (sets average or peak iris automatically)
|
|
Structure emphasis
|
Function to adjust the sharpness of the subject structure. SE (Structure Emphasis) 4 level, DH (fine section) -4~+9, DL (structure section) -4~+9.
|
|
Tone
|
Function to emphasize slight differences between colors by emphasizing the degree of vividness of color. ON/OFF.
|
|
Enlargement of the image
|
Function to enlarge the endoscopic image.
|
|
Special light observation mode
|
BLI, BLI-bright, LCI
|
|
FICE
|
Ten settings available.
|
|
Mask types
|
Type 1, Type2, Type 2/Dual Mode.
|
|
Freeze mode
|
Function to freeze the endoscopic images.
|
|
Peak detection
|
Function to obtain the highest contrast image.
|
|
Shutter speed
|
Normal 1/60-1/200, High 1/100-1/400, High (zoom scope) 1/100-1/800
|
|
Assignment of switches
|
Scope Switch (1-5), Multi buttons on the front panel (1.2), Foot Switch (1,2).*1
|
|
Other functions
|
Electronic Zoom, PoP Function, Network function, Dual Mode function.
|
Applicable endoscope
|
|
700 system scopes, 600 system scopes, 500 system scopes
|
Data display
|
Remote control
|
Fujifilm specified peripherals can be controlled.
|
|
Patient information
|
Patient ID, Patient Name, Sex, Age, Date of Birth, Comments, Hospital name, Doctor name *2
|
|
Other information
|
Timer, Laptime
|
|
Recording status
|
Digital printer status, shooting counter, number of recordable images in internal storage device
|
|
Image quality setting status
|
Structure emphasis, Tone, Electronic Zoom Ratio, IEE observation modes, Focus Indicator.
|
Image recording
|
Image compression rate
|
TIFF: no compression, JPEG: approx. 1/5 , 1/10, 1/20
|
|
Number of recordable images in internal storage device
|
TIFF: 840, JPEG 1/20: 21,690, JPEG 1/10: 16,270, JPEG 1/5: 5,910 *
|
|
Recommended external storage device
|
Swissbit SFU-22048 E1BP2TO-I-MS-111-STD or SFU22048E3BP2TO-I-MS-121-STD
|
|
Searching and displaying images
|
Search screen: Inspection No., Patient ID, Date of Inspection. Display: List, Thumbnail, Enlargement.
|
Data presetting
|
Doctors’ name
|
Up to 20 doctors’ names.
|
|
Setting by doctor
|
The information such as color tone, iris mode, contrast, brightness, IEE observation modes are kept by setting the doctor’s name.
|
|
Clinical procedure
|
Up to 20 procedures.
|
Memory backup
|
When using lithium battery
|
6 years (based on FUJIFILM criteria)
|
Control connector
|
|
Light source: 1, Remote: 2, Peripherals: 2, Keyboard: 1, Card reader: 1, Digital printer: 1, Footswitch: 1, Network: 1.
|
Category of medical electric equipment
|
Type of protection against electric shock
|
Class I equipment
|
|
Degree of protection against electric shock
|
Type BF applied part
|
|
Degree of explosion protection
|
Prohibited in oxygen-rich environment/ flammable gas atmosphere.
|
|
|
BL-7000
|
Power
|
|
100 to 240 V ~ ± 10 %, 50/60 Hz
|
Power consumption (rated)
|
|
1.2-0.7 A
|
Dimensions (W×H×D)
|
|
390×155×485 mm (incl.projection)
|
Weight
|
|
12Kg
|
Illumination
|
Illumination source
|
LED, qualifies 300W Xenon lamp intensity
|
|
Durability of LED
|
6 years (based on FUJIFILM criteria)
|
|
Lighting system
|
Switching regulator
|
|
Light control method
|
LED Auto power control
|
|
Light cooling method
|
Forced air cooling
|
|
Special light observation mode
|
BLI, BLI-bright, LCI
|
|
Maximum light output
|
1400 lm (based on FUJIFILM criteria)
|
|
Maximum air supply pressure
|
65 kPa
|
Automatic brightness adjustment
|
Automatic brightness adjustment method
|
Brightness is automatically adjusted according to the video output (manually possible).
|
Air supply
|
Pump
|
Diaphragm method pump
|
|
Air supply pump
|
HI/MID/LOW/OFF
|
Water supply
|
Method
|
Feeds water by pressurizing the detachable water container with air.
|
Indicators on front panel
|
Transmitted illumination
|
The light flashes with the maximum light intensity. Used to check the position of the distal end from outside the body.
|
|
Light limitation
|
To avoid the blood of a bleeding patient becoming clotted by the illuminating light. Used to limit the maximum light intensity.
|
|
Illumination mode
|
OFF/1/2/3. Observation modes can be switched by pressing the illumination mode button.
|
Memory of set value
|
|
Set values are maintained even after turning off the system.
|
Category of medical electric equipment
|
Type of protection against electric shock
|
Class I equipment
|
|
Degree of protection against electric shock
|
Type BF applied part
|
|
Degree of explosion protection
|
Prohibited in oxygen-rich environment/ flammable gas atmosphere.
|